GD-300MT Vicat Softening Point Tester is used to determine the HDT and VICAT softening point temperature of non-metal materials such as plastic, rubber, nylon, and electrically insulating materials. It is widely used in the chemical industry, R&D departments, and academies or colleges.
Availability:
Quantity:
HTD & Vicat Softening Temperature Tester, ISO 75, ASTM D648
  • GD-300MT

  • GOLD

GD-300MT  HTD & Vicat Softening Temperature Tester

  Introduction:  

GD-300MT Vicat Softening Point Tester is used to determine the HDT and VICAT softening point temperature of non-metal materials such as plastic, rubber, nylon, and electrically insulating materials. It is widely used in the chemical industry, R&D departments, and academies or colleges.

The machine uses an operating interface to monitor the test temperature in real-time, draw the curve of temperature and deformation in real-time, print the test report, and realize any point of the historical disk, mouse click deformation, and temperature curve. The interface automatically displays the amount of deformation and temperature values at the site. A high-precision displacement sensor(intelligent telemeter) is used to measure deformation.

Determination of temperature at softening point of thermoplastic Vicat: The temperature at which a standard pressure needle is inserted into the surface of a thermoplastic specimen at a depth of 1mm under a load condition is determined when the temperature is heated at a uniform rate.

Determination of thermal deformation temperature of plastic bending load: Determination of plastic specimen immersed in a suitable liquid heat transfer medium at an equal rate of warming. Under the action of a simple beam-type static bending load, the specimen bending deformation reaches the specified temperature.

Ultra intelligent design

After the input parameters and the sample are placed, the test is started. The meter will automatically clear zero. The instrument automatically balances the temperature and automatically begins to heat up. After the last sample is measured, the experiment will be automatically stopped and the data will be automatically recorded.



  Standards:  

ISO 75-1 Plastics - Determination of temperature of deflection under load  

Part 1: General test method

Part 2: Plastics and ebonite

ISO 36 Plastics – Thermoplastic materials – Determination of Vicat softening temperature (VST)

ASTM D648 Standard Test Method for Deflection Temperature of Plastic Under Flexural Load in the Edgewise Position

Note: It also conforms to other equivalent standards.



  Features:  

  • Advanced control system:It has high stability and is easy to install, test and analyze.

  • User friendly software control system:The temperature and displacement can be displayed by imitating digital tube, and the test curve can be drawn. It provides a variety of coordinate curves for centralized data management. It can analyze the test data, print the test curve, test data and test report.

  • Four kinds of safety protection:Multiple software and hardware security designs, such as mechanical temperature upper limit protection switch, software temperature upper limit protection, automatic test stop and administrator operation key, ensure the safe and stable operation of the test.

  • High precision measurement and control system:

  • High performance temperature sensor and advanced capacitive grating digital dial indicator are used to accurately measure the small changes of the sample. Advanced PID algorithm and full photoelectric isolation phase modulation module are used to control the temperature rise, so as to ensure that the temperature rise rate error is less than 0.5 ℃.



  Technical Parameters:  

1. Rising temperature range: Ambient temperature~300℃;

2. Constant temperature control range: 50℃~200℃

3. Heating rate: (120±10)℃/h  [(12.0±1.0)℃/6min]

(50±5)℃/h  [(5.0±0.5)℃/6min]                

4. Max temperature display error: ±0.5℃

5. Max deflection display error:±0.005 mm

6. Deflection measure range: 0.001 mm~10.0 mm

7. Test stations: 3

8. Temperature sensors: 1

9. Weight of loading rod and load-carrying plate: 64g

10. Heating medium: Silicone oil or transformer oil (The recommended viscosity of Silicon oil is 100Cst~200Cst, and its flash point should higher than 300℃; the transformer oil should reach 240℃)

11. Cooling method: natural cooling(above 150℃); natural cooling or water cooling (below 150℃)

12. Heating power: 4kW

Aggregate power capacity: 6kW

13. Test frame lifting method: Manual



   Welcome to contact us:  

Contact Person: Kiana Feng

Whatsapp/Wechat: +86 15922982738

Mobile: +86 15922982738

Fax: +86-23-62940030

Email: gold20@hy-industry.com


Previous: 
Next: 
Send Us A Message

  Site Map

Contact Us

  25F, China Resources Building, Xiejiawan, Jiulongpo DistrictChongqing,China
  +86-23-62984892
  +86-15922982738
   
gold20@hy-industry.com
  +86-23-62940030

Quick Links

Send Us A Message

​Copyright © 2018  Chongqing Gold Mechanical & Electrical Equipment Co.,Ltd. All Rights Reserved.